» Topics » Advice »Anti-decubitus compressor ARIES DX-2 device

Anti-decubitus compressor ARIES DX-2 device


This compressor got into repair, it shakes badly. Perhaps someone will come in handy the device of this unit

Disassemble

The compressor itself




The distributor mounted on a stepper motor, for distributing air alternately into two channels




We disassemble the distributor


It is necessary to press down the spring to release the stopper





Air comes from the compressor, then with the help of a stepper motor it distributes the direction: either to the right or to the left tube (pumps two separate halves of the mattress

All whole in working condition, hoses without gusts.
One option remains, tighten the control valve


Putting it back

Like all!

Collections: Compressor

Suitable for topic

Related topics

Add a comment

    • smilesmilesxaxaokdontknowyahoonea
      bossscratchfoolyesyes-yesaggressivesecret
      sorrydancedance2dance3pardonhelpdrinks
      stopfriendsgoodgoodgoodwhistleswoontongue
      smokeclappingcraydeclarederisivedon-t_mentiondownload
      heatirefullaugh1mdameetingmoskingnegative
      not_ipopcornpunishreadscarescaressearch
      tauntthank_youthisto_clueumnikacuteagree
      badbeeeblack_eyeblum3blushboastboredom
      censoredpleasantrysecret2threatenvictoryyusun_bespectacled
      shokrespektlolprevedwelcomekrutoyya_za
      ya_dobryihelperne_huliganne_othodifludbanclose
17 comments
Guest Maxim
where do the pet stores or bird markets?
Despite the fact that judging by the photo, this is a compressor for the aquarium, adapted for the mattress! smile
Guest Maxim
But I didn’t understand a bit, what about pet stores or bird markets? How to find this item there?
Or maybe there is a name for this part, or maybe something like that can be ordered from Ali?
Guest Maxim
Tell me, can you know where you can find this part in the compressor?
I would look at pet stores, bird markets, or a mattress seller. Recently, I myself bought parts for an electronic blood pressure monitor, at the pharmacy we gave the address of the service center! yes
Guest Maxim
Good afternoon. Tell me, can you know where you can find this part in the compressor? She broke, and the pump does not work well. Here's how you got a round rubber black valve:
Good evening, there was such a question, when the compressor emits a rhythmic knock, it’s quiet and kind of not annoying, but I decided to take it apart and check the reason in the distributor, here’s the question, it’s like ceramic material is like a water tap, and so these rubbing parts stick to each other and this is the reason for knocking, the surfaces seem to be slippery but when touched with each other they directly stick like magnetized ones, if you can advise something in this case, can they be lubricated, or have already worn out if you not hard tell me solved Ie.
Author
Remove the adjustment knob and tighten the valve a little, (with pliers), to the right side ... Do not overdo it ...
Guest KIRILL
Please tell us in more detail how to tighten the control valve of such a compressor.
And how to tighten this valve? I already have 5 such compressors, and everyone has the same problem. After some time, they stop pumping normally. Thank you.
What a twist.
On the site, the residents of our sites are discussing the repair of medical equipment. Even if simple.
As a medical technician, experienced in his field, I will also insert a couple of cents.
The design of budget compressors for decubitus mattresses is not diverse. But some models have filters at the air inlet to the compressor. And they tend to lose bandwidth. It was possible to throw out a thick felt and cut circles from a disposable mask into several layers. The switching speed of the air channels does not change. Pump performance only.
As noted above, if the pump fails, it is perfectly acceptable to select the power from the aquarium.
Thanks so much for this article! Specifically registered to thank. In the photographs, you so clearly explained that I immediately became bolder and repaired the Meditek compressor with my own hands. He lay idle for about 5 years and began to pump both sections equally and remained without any cycles of alternating blowing-pumping. Having disassembled the distributor, I immediately realized that the lubricant inside it had dried / lost its properties. I used a thick silicone grease that came with the Atoll drinking water filter. Now it shakes like a clock! By the way, there are almost no materials on this topic. Thanks again.
Yes, actually, and not at all.
Korolev,
THANKS.
Remembered! In pet stores, air flow regulators are sold to the aquarium, try them.
I think that we should try to play with the diameter of the exhaust pipe, and the cycle time of 12 minutes is set by the revolutions of the engine - gearbox, here it will be more difficult to adjust.
Thanks for the answer. Is it possible to do something that would not lower the end?
I think the performance of this particular compressor is not enough for 6 minutes to pump the required volume. I used to have exactly such a barely barely air in the small aquarium.
Good afternoon. I have a question. How to modify the compressor so that one cycle of work is not 12 minutes but for example 6. To inflate and zduf cells more often and even when one row of cells is deflated and another is inflated, so that the suf is not completely finished. Thanks.

We advise you to read: